Cell Sorting 101
Cell sorting is a specialized flow cytometry technique used to physically separate selected cells from a mixed sample.
A standard flow cytometer measures the physical and fluorescent characteristics of cells but normally sends all analyzed cells to waste. A cell sorter analyzes those same characteristics and then directs selected cells into collection tubes, plates, or other containers.
Researchers can sort cells based on:
• Cell size
• Internal complexity
• Surface-marker expression
• Fluorescent protein expression
• Cell viability
• DNA content
• Functional fluorescent probes
• Combinations of multiple characteristics
The isolated cells can then be used for cell culture, microscopy, molecular analysis, functional assays, sequencing, cloning, or other downstream applications.
Successful cell sorting requires more than identifying a population on a flow cytometry plot. Sample preparation, cell viability, panel design, instrument settings, gating, collection conditions, and post-sort handling can all affect the final purity and condition of the sorted cells.
What Is Cell Sorting?
Cell sorting is the physical isolation of one or more cell populations from a mixed sample.
In fluorescence-based cell sorting, cells pass individually through the instrument’s lasers. The cytometer measures their light-scatter and fluorescence characteristics and determines whether each cell matches the sorting gates created by the operator.
In many droplet-based sorters, the liquid stream is broken into droplets after the cells pass through the lasers. Droplets containing selected cells receive an electrical charge and are deflected into designated collection containers. Unselected cells are directed to waste or another collection container.
Cell sorting may be used to isolate:
• A broad immune-cell population
• A rare cell subset
• Live transfected cells
• Fluorescent reporter-positive cells
• Stem or progenitor cells
• Tumor cells
• Antigen-specific lymphocytes
• Cells at a particular stage of the cell cycle
• Single cells for cloning or sequencing
Why Use Cell Sorting?
- To Isolate a Specific Cell Population
Biological samples often contain several cell types.
For example, peripheral blood may contain T cells, B cells, NK cells, monocytes, granulocytes, platelets, and other populations. Cell sorting allows researchers to isolate one selected population from that mixture.
A researcher may sort:
• CD3-positive T cells
• CD19-positive B cells
• CD14-positive monocytes
• CD56-positive NK cells
• CD4-positive regulatory T cells
• A specific memory T-cell population
• A fluorescently labeled tumor-cell population
The selected markers should be based on the biological identity of the target population rather than on one marker alone when several cell types may share that marker.
- To Obtain a More Purified Sample
Mixed cell populations can make molecular and functional results difficult to interpret.
If RNA is extracted from an unsorted sample, the measured gene expression represents all cells in that sample. Sorting the target population first allows the analysis to focus more specifically on that population.
Cell sorting can improve the interpretation of:
• Gene-expression studies
• Protein analysis
• Cell-culture experiments
• Functional assays
• Cytotoxicity experiments
• Receptor studies
• Drug-response experiments
• Single-cell sequencing
- To Study Rare Cells
Some cell populations represent only a small percentage of the starting sample.
Flow cytometry can identify these rare populations, and a cell sorter can physically collect them for further study.
Examples may include:
• Antigen-specific T cells
• Circulating tumor cells
• Stem-cell populations
• Rare progenitor cells
• Minimal residual disease populations
• Small activated immune-cell subsets
Rare-cell sorting requires enough starting cells, accurate gates, careful exclusion of artifacts, and sufficient collection time.
- To Produce Clonal Cell Populations
A sorter can place one cell into each well of a multiwell plate.
This is known as single-cell sorting or index sorting, depending on the instrument and the information recorded.
Single-cell sorting may be used for:
• Monoclonal cell-line development
• Antibody-producing cell isolation
• CRISPR-edited clone selection
• Hybridoma screening
• Single-cell sequencing
• Functional testing of individual cells
The success of single-cell cloning depends on both accurate cell deposition and the ability of an individual sorted cell to survive and grow.
- To Recover Living Cells
Cells can be sorted alive when the panel uses cell-surface markers, fluorescent proteins, or live-cell-compatible probes.
Live sorted cells may be used for:
• Cell culture
• Differentiation experiments
• Functional testing
• Adoptive-transfer studies
• Drug screening
• Microscopy
• RNA or DNA analysis
Fixation and permeabilization may be used when intracellular markers are required, but fixed cells cannot be returned to normal live-cell culture.
Cell Sorting Versus Magnetic Cell Separation
Flow cytometry cell sorting and magnetic cell separation can both enrich a target population, but they provide different levels of control.
- Flow Cytometry Cell Sorting
Flow cytometry sorting can isolate cells based on several markers at the same time.
It can also separate populations based on:
• Different levels of marker expression
• Positive and negative marker combinations
• Fluorescent reporter intensity
• Viability
• Light-scatter characteristics
• Multiple sequential gates
This makes it useful for isolating highly specific populations.
- Magnetic Cell Separation
Magnetic separation uses antibody-coated magnetic particles to enrich or remove cells expressing a selected surface marker.
It is often faster and easier for processing large samples, but it usually provides less multiparameter control than flow cytometry sorting.
Magnetic enrichment may be performed before flow sorting when the target population is extremely rare. Removing large numbers of unwanted cells can shorten the final flow-sorting time.
How Does a Flow Cytometry Cell Sorter Work?
A typical sorting process includes several connected steps.
First, the sample enters the fluidics system and is focused so that cells pass through the lasers individually.
The instrument measures forward scatter, side scatter, and fluorescence signals for each event.
The software compares those measurements with the sorting gates selected by the operator.
If a cell matches a gate, the sorter determines which droplet will contain that cell after the stream breaks into droplets.
That droplet is given an electrical charge.
Charged droplets are then deflected toward the appropriate collection container. Uncharged or unselected droplets continue toward waste.
Accurate sorting depends on stable fluidics, proper stream alignment, correct drop-delay timing, appropriate controls, and a clearly defined target population.
Step 1: Define the Goal of the Sort
Before preparing the cells, researchers should determine exactly what the sorted population will be used for.
Important questions include:
• Which population needs to be isolated?
• How many sorted cells are required?
• Does the population need to remain alive?
• Is maximum purity or maximum recovery more important?
• Will the cells be cultured after sorting?
• Will the cells be used for RNA, DNA, or protein analysis?
• Is bulk sorting or single-cell sorting required?
• How rare is the target population?
• Which markers are needed to identify it?
The downstream application affects nearly every part of the workflow.
Cells intended for culture require sterile handling and a supportive collection medium.
Cells intended for RNA analysis require rapid processing and protection from RNA degradation.
Cells intended for single-cell sequencing may need to be deposited directly into a lysis buffer or a specific collection plate.
Step 2: Select a Sorting Strategy
A sorting strategy defines how the target cells will be identified and collected.
Positive Selection
Positive selection collects cells that express a desired marker.
For example:
• CD3-positive T cells
• CD19-positive B cells
• GFP-positive transfected cells
• CD34-positive progenitor cells
Positive selection is often straightforward when the target marker is specific and clearly separated from the negative population.
Negative Selection
Negative selection removes cells expressing unwanted markers and collects cells lacking those markers.
This may be useful when labeling the target population could alter its function or when the target lacks one unique identifying marker.
Combination Gating
Many populations require a combination of positive and negative markers.
For example, a target population may be defined as:
• Live
• Singlet
• CD45 positive
• CD3 positive
• CD4 positive
• CD25 high
• CD127 low
Combination gating is more specific than sorting on one marker alone.
Reporter-Based Sorting
Cells expressing GFP, mCherry, or another fluorescent protein can be sorted without fluorescent antibodies.
An appropriate nonfluorescent control is still required to establish the negative range and account for cellular autofluorescence.
Step 3: Design the Sorting Panel
A sorting panel should identify the target population while preserving cell health and minimizing unnecessary complexity.
The panel may include:
• A viability dye
• Lineage markers
• Exclusion markers
• Target-population markers
• Activation or differentiation markers
• A fluorescent reporter
The number of markers should be based on what is necessary to identify the population accurately.
Adding unnecessary markers can increase:
• Staining time
• Sample handling
• Spectral overlap
• Spreading error
• Reagent cost
• Gating complexity
Antibodies should be titrated under the actual sorting conditions.
Using too much antibody can increase background staining and spreading. Using too little may make positive and negative cells difficult to separate.
Step 4: Prepare a High-Quality Single-Cell Suspension
Cell sorters require cells to pass through the instrument individually.
Cell clumps can cause:
• Nozzle blockages
• Unstable sample flow
• Increased doublets
• Poor sorting accuracy
• Lower cell recovery
• Interrupted sorts
Suspension cells should be gently mixed before staining and sorting.
Adherent cells and solid tissues must be dissociated into a single-cell suspension. The dissociation method should be strong enough to release the cells but gentle enough to preserve viability and important surface markers.
The preparation should remove:
• Large tissue fragments
• Cell aggregates
• Excessive debris
• Undissolved material
• Unwanted red blood cells when appropriate
Filtering the final sample through an appropriate cell strainer shortly before the sort can reduce clumps. The filter size should be selected according to the size of the cells and the sorter’s nozzle.
Step 5: Count the Cells and Measure Viability
The sample should be counted before staining and again before sorting when substantial cell loss may have occurred.
Accurate cell counts help determine:
• Whether enough starting cells are available
• How much antibody is required
• The final sample concentration
• The approximate sorting time
• Whether enough target cells can realistically be recovered
Viability is especially important because dead cells can increase debris, nonspecific antibody binding, autofluorescence, and aggregate formation.
A high total cell count does not guarantee that enough usable cells are present. The viable cell count is often more informative when planning a live-cell sort.
Step 6: Stain the Cells
A typical live-cell surface-staining workflow includes:
- Preparing the required controls.
- Aliquoting the cells.
- Applying a blocking step when needed.
- Adding the antibody panel.
- Incubating under validated conditions.
- Washing away unbound antibody.
- Resuspending the cells in a sorter-compatible buffer.
- Adding a viability dye according to its instructions.
- Filtering the sample before loading it onto the sorter.
Cells should be protected from unnecessary light exposure, especially when the panel contains light-sensitive fluorophores or tandem dyes.
The staining buffer should maintain cell viability while reducing nonspecific binding and cell aggregation.
The exact buffer should be compatible with:
• The cell type
• The antibodies
• The viability dye
• The instrument
• The downstream application
Researchers should consult the sorting facility before adding unfamiliar substances because some additives can affect fluidics, fluorescence, or instrument cleaning.
Step 7: Prepare the Necessary Controls
Controls are needed to correctly configure the instrument and define the sorting gates.
Unstained Control
The unstained control shows the sample’s natural autofluorescence and helps identify the negative fluorescence range.
Single-Stained Controls
Single-stained controls are used to calculate compensation on conventional flow cytometers or spectral-reference signatures on spectral sorters.
A separate control is generally required for each fluorophore.
Controls should be sufficiently bright and should match the fluorophores used in the experiment.
Fluorescence-Minus-One Controls
A fluorescence-minus-one control contains every fluorophore in the panel except one.
It can help define a gate when a marker is:
• Dim
• Continuously expressed
• Affected by fluorescence spread
• Difficult to separate into positive and negative groups
An FMO is not automatically required for every marker. It is most useful when the gate cannot be defined confidently using biological controls.
Viability Controls
A positive dead-cell control can help establish the viability-dye-positive range.
Biological Controls
Biological controls may include:
• A known negative population
• A known positive population
• Untreated cells
• Stimulated cells
• Untransfected cells
• A fluorescence-reporter-negative cell line
Biological controls can be more useful than relying only on arbitrary gates.
Step 8: Prepare the Collection Containers
Sorted cells should be collected into a solution that supports the intended downstream application.
Possible collection containers include:
• Tubes
• Multiwell plates
• Microcentrifuge tubes
• Culture plates
• PCR plates
• Single-cell sequencing plates
Collection media may include:
• Cell-culture medium
• A protein-containing buffer
• Serum-containing medium
• RNA-preservation solution
• A lysis buffer
• A downstream-assay-specific solution
For live-cell collection, tubes may be prepared with a small volume of supportive medium to prevent cells from landing in an empty, dry container.
Collection conditions should be selected based on:
• Cell sensitivity
• Sort duration
• Required sterility
• Final cell concentration
• The next experimental step
Step 9: Complete Instrument Setup and Quality Control
Cell sorters should be operated by trained personnel following the instrument-specific standard operating procedure.
Before sorting, the operator typically evaluates:
• Fluid levels
• Waste levels
• Laser performance
• Detector performance
• Stream stability
• Nozzle condition
• Droplet formation
• Drop-delay accuracy
• Deflection alignment
• Aerosol containment
• Collection-tube alignment
Quality-control beads may be used to confirm that the instrument is performing within the expected range.
Sorters are complex instruments, and droplet-based sorting can produce aerosols. Samples must be evaluated through the facility’s biosafety process, particularly when they contain infectious, potentially infectious, genetically modified, or human-derived material.
Step 10: Select an Appropriate Nozzle and Pressure
The sorter’s nozzle size and operating pressure influence sorting speed and cell stress.
A smaller nozzle may support faster sorting but may expose cells to greater pressure and may be inappropriate for large or fragile cells.
A larger nozzle may be more suitable for:
• Large cells
• Fragile cells
• Primary cells
• Stem cells
• Dissociated tissue cells
• Cells that are easily damaged
The nozzle must be large enough to allow cells to pass without clogging.
There is no single nozzle size that is best for every sample. The correct choice depends on the instrument, cell diameter, cell fragility, required speed, and downstream application.
The operator and researcher should discuss these requirements before the sort begins.
Step 11: Acquire a Small Sample and Establish the Gates
Before activating the sorting mechanism, a portion of the sample should be analyzed.
A typical sorting gate may include:
- Time Gate
A time plot can reveal temporary changes in flow rate, pressure, fluorescence, or stream stability.
Events collected during an unstable period may need to be excluded.
- Cell Gate
Forward and side scatter can help separate cells from obvious debris.
This gate should not be so restrictive that it removes the target cells, especially if the target population has unusual scatter characteristics.
- Singlet Gate
Doublets and aggregates should be excluded using pulse measurements such as area, height, or width.
A doublet containing one target cell and one unwanted cell may appear positive for both populations and reduce sort purity.
- Viability Gate
Dead cells should be excluded during live-cell sorting.
- Lineage Gate
Lineage markers can identify the broad cell group containing the target population.
- Final Target Gate
The final gate should identify the desired population using the required marker combination.
The gate should be based on controls and biological expectations rather than being moved simply to collect more cells.
Step 12: Choose the Sort Mode
Most sorters offer different modes that balance purity, yield, and recovery.
The exact names vary by instrument, but common priorities include:
Purity-Focused Sorting
A purity-focused mode rejects uncertain droplets that may contain both a target cell and an unwanted cell.
This can improve the purity of the collected population but may reduce the total number of target cells recovered.
Yield-Focused Sorting
A yield-focused mode attempts to recover more target cells.
It may accept some droplets that have a higher risk of contamination, potentially lowering purity.
Single-Cell Sorting
Single-cell modes are designed to deposit one selected cell into each designated well.
The sorter must be aligned accurately with the plate, and deposition should be verified according to the facility’s procedure.
The selected mode should be based on the downstream experiment.
For example, a molecular assay requiring a very pure population may prioritize purity.
An experiment involving an extremely rare population may prioritize recovery.
Step 13: Perform the Sort
Once the instrument, sample, controls, and gates have been reviewed, the sort can begin.
During the sort, the operator should monitor:
• Event rate
• Stream stability
• Sample flow
• Sorting efficiency
• Abort rate
• Collection-container position
• Cell aggregation
• Sort-gate frequency
• Temperature when temperature control is available
Samples should be mixed gently as needed to prevent cells from settling.
The event rate should not be increased so far that coincident events, aborts, or sorting errors reduce the quality of the collection.
Faster is not always better. A slower, stable sort may provide better purity, viability, and recovery than an unstable high-speed sort.
Step 14: Protect the Cells During Long Sorts
Long sorting procedures can expose cells to:
• Nutrient limitations
• Temperature changes
• Light exposure
• Mechanical stress
• Concentration changes
• Settling
• Aggregation
• Loss of viability
Sensitive samples may benefit from temperature-controlled sample and collection holders when appropriate.
The sample should be monitored for new clumps, especially during a long sort.
Large collection volumes may need to be concentrated after sorting, but every additional centrifugation or transfer step may cause cell loss.
Step 15: Evaluate Post-Sort Purity
A small portion of the collected cells can be run back through the analyzer or sorter.
This is called a post-sort purity check.
The same general gating strategy used during sorting should be applied to the post-sort sample.
The purity calculation is commonly based on the percentage of collected events that still fall within the target gate.
A post-sort check can identify:
• Contamination by unwanted cells
• Improper gate placement
• Doublet contamination
• Sorting instability
• Collection errors
• Changes in marker expression
A high software-reported sorting efficiency does not replace a biological post-sort purity check.
Step 16: Measure Post-Sort Viability and Recovery
Purity is only one measure of sorting success.
Researchers should also consider:
- Viability
What percentage of the collected cells remain alive?
A sample can be highly pure but unsuitable for culture if most of the cells are damaged.
- Recovery
How many of the expected target cells were actually collected?
Recovery may be lower than expected because of:
• Cell loss during preparation
• Cell settling
• Aggregation
• Conservative gates
• Sorting aborts
• Cells remaining in the tubing
• Transfer loss
• Post-sort centrifugation loss
- Yield
How many usable cells were obtained?
The word “yield” is sometimes used differently between laboratories, so the calculation and meaning should be clearly documented.
- Function
Do the sorted cells still perform the required biological function?
Depending on the study, this may be evaluated through:
• Cell growth
• Cytokine production
• Cytotoxicity
• Differentiation
• Phagocytosis
• Receptor signaling
• Gene expression
Bulk Sorting Versus Single-Cell Sorting
Bulk Sorting
Bulk sorting collects many cells with the same phenotype into one container.
It is useful for:
• Cell culture
• RNA extraction
• Protein analysis
• Functional experiments
• Population-level sequencing
• Drug-response studies
Single-Cell Sorting
Single-cell sorting deposits individual cells into separate wells.
It is useful for:
• Clonal expansion
• Single-cell sequencing
• Antibody discovery
• Hybridoma screening
• CRISPR clone isolation
• Genotype-to-phenotype studies
Single-cell sorting requires accurate deposition, careful plate setup, and conditions that allow one cell to survive in a well by itself.
How to Sort Rare Cell Populations
Rare-cell sorting requires additional planning.
First, the expected target frequency should be estimated.
For example, if the population represents 0.1% of the viable sample, approximately one target event may be present per 1,000 viable events before accounting for processing losses.
Researchers should consider:
• Starting sample size
• Expected target frequency
• Required final cell number
• Pre-sort viability
• Sort efficiency
• Expected recovery
• Total sorting time
Magnetic pre-enrichment may reduce the number of irrelevant cells that must pass through the sorter.
A dump channel may also be used to combine several unwanted markers into one fluorescence parameter, allowing those populations to be excluded efficiently.
Rare-population gates should be validated carefully because debris, dead cells, doublets, nonspecific staining, and fluorescence spread can resemble rare positive events.
Purity, Recovery, and Viability: Understanding the Tradeoffs
The highest purity, highest recovery, and highest viability may not always be achieved simultaneously.
A restrictive gate may produce a highly pure population but exclude valid target cells.
A broad gate may increase recovery but include contaminating cells.
A high event rate may shorten the sort but increase coincident events and sorting aborts.
A smaller nozzle may increase speed but may be more stressful for fragile cells.
The best sorting conditions depend on the experiment’s main priority.
Researchers should decide before the sort whether the most important outcome is:
• Maximum purity
• Maximum number of cells
• Maximum viability
• Minimum sorting time
• Accurate single-cell deposition
• Preservation of cell function
Common Cell-Sorting Problems
- Nozzle Blockages
Possible causes include:
• Cell clumps
• Large cells
• Tissue fragments
• Excessive cell concentration
• Inadequate filtering
• Precipitated reagents
The sample should be filtered, maintained as a single-cell suspension, and prepared at a concentration recommended for the instrument.
- Low Cell Viability
Possible causes include:
• Harsh tissue dissociation
• Long staining procedures
• Incorrect buffer conditions
• Excessive centrifugation
• Long sort times
• High pressure
• Inappropriate temperature
• Toxic staining reagents
Viability should be measured before the sort so that damage caused during sorting can be distinguished from poor starting-sample quality.
- Low Purity
Possible causes include:
• Poorly defined gates
• Doublets
• Inadequate fluorescence separation
• Excessive event rate
• Incorrect drop-delay settings
• Unstable stream
• Improper collection alignment
- Low Recovery
Possible causes include:
• A target population rarer than expected
• Cell loss during staining
• Cell settling
• Sample remaining in the tube
• High abort rates
• A restrictive sort mode
• Transfer loss after collection
• Cell death after sorting
- Cells Do Not Grow After Sorting
Possible causes include:
• Sorting stress
• Low post-sort viability
• Inappropriate collection medium
• Insufficient cell density
• Lack of required growth factors
• Damage from staining reagents
• Single cells requiring feeder cells or conditioned medium
- Unexpected Fluorescence
Possible causes include:
• Autofluorescence
• Nonspecific antibody binding
• Dead-cell staining
• Incorrect compensation or unmixing
• Tandem-dye degradation
• Fluorescent-protein spillover
• Inappropriate controls
Biosafety Considerations for Cell Sorting
Droplet-based cell sorting can generate aerosols, especially when the stream becomes unstable or the nozzle is obstructed.
Samples should be reviewed according to:
• The source of the cells
• Potential infectious agents
• Human or animal origin
• Genetic modification
• Viral transduction
• Fixation status
• Known or unknown pathogens
The sorting facility should determine the required containment, personal protective equipment, aerosol-management procedures, disinfection method, and operator training.
Researchers should never assume that a sample is safe to sort simply because it is handled under BSL-2 conditions elsewhere in the laboratory.
The sorter’s biosafety requirements may be more restrictive because of aerosol generation.
Best Practices for Flow Cytometry Cell Sorting
To improve sorting quality:
- Define the target population and downstream application before designing the panel.
- Confirm that the sorter has the required lasers and detectors.
- Discuss cell size, cell type, biosafety, and collection requirements with the sorting facility.
- Prepare a clean single-cell suspension.
- Remove clumps and large debris.
- Count the cells and measure viability.
- Titrate antibodies under the actual sorting conditions.
- Use a viability dye for live-cell sorting.
- Include appropriate single-stained controls.
- Use biological and FMO controls when needed to define difficult gates.
- Select fluorophores that clearly separate the target population.
- Filter the sample shortly before sorting.
- Choose a nozzle and pressure appropriate for the cells.
- Exclude debris, doublets, and dead cells.
- Select the sort mode based on purity, yield, or single-cell requirements.
- Avoid unnecessarily high event rates.
- Collect cells into an appropriate medium or assay solution.
- Monitor the sample and stream throughout the sort.
- Perform a post-sort purity check.
- Measure post-sort viability and cell recovery.
- Process or culture the sorted cells promptly.
- Document the complete sorting and gating strategy.
Producing Reliable Cell-Sorting Results
Cell sorting allows researchers to isolate precisely defined populations from complex biological samples.
The process begins before the sample is placed on the instrument. Researchers must define the target population, design an appropriate panel, prepare a viable single-cell suspension, establish controls, and select collection conditions that support the downstream application.
During sorting, the operator must maintain stable fluidics, accurate droplet timing, appropriate event rates, and clearly defined gates.
After sorting, purity, viability, recovery, and cell function should be evaluated. A successful sort is not simply one that collects cells. It is one that produces enough cells of the correct identity and quality to support the next experiment.
When sample preparation, staining, instrument setup, gating, biosafety, and post-sort handling are carefully controlled, flow cytometry cell sorting can provide highly purified cell populations for culture, sequencing, functional testing, and other advanced research applications.